Health Tag Respiratory Rate Validation

Health Tag Respiratory Rate Validation

Objective:
The objective of the study was to conduct a respiratory rate (RR) accuracy evaluation comparing the Health Tag to an FDA-cleared end-tidal carbon dioxide monitor (etCO2) reference device in a mixed population of 20 adults, ranging from healthy to those with varying cardiopulmonary diseases.

Reference device:
The reference etCO2 device used in this study was the GE Healthcare S5Multi-parameter Monitor™, M-COVX module.

Results:
  • The range evaluated during the study was 8-35 breaths per minute (bpm).
  • The accuracy of the Device Under Test (DUT), Health Tag, passed with an average root mean square (ARMS) value of 0.8 bpm.
Health Tag respiratory rate vs reference etCO2 respiratory rate

To access the full validation study, please contact us at RPM@spirehealth.com